OI Analytical FS3700 Continuous Flow Analyzer
Automates Wet Chemistry Test Procedures
- Can perform both Segmented Flow Analysis (SFA) and Flow Injection Analysis (FIA)
- Validates Hardware Set Up and Performance for every method supplied
- Multiple FS3700’s can be linked to provide additional channels for simultaneous analysis.
- Modular, Flexible Hardware
Versatile
Featuring an unmatched modular design, the FS3700 allows for flexible and versatile functionality to help meet your lab’s goals. Run different flow methods in different channels in tandem on the FS3700. Add pre-configured chemistry cartridges and perform validated analysis methods. Link Multiple systems together to provide additional analytical channels.
Easy Validation
Hardware configurations and performance of each supplied method are validated to provide users with a complete analysis solution. Validated methods for aqueous samples, soil, and plant extracts that support environmental compliance monitoring, process optimization, and research applications are available in easy-to-use, pre-configured chemistry cartridges.
Simple-to-Learn Interface
FlowView™ Software is unmatched in providing a user-friendly, icon based interface that simplifies scheduling, operation, and report generation from the FS3700. Compatible with Windows 32- or 64- bit operating systems, it requires minimal training and makes exporting and retrieving data a breeze.
FlowView™ features:
- On-the-fly sequence editing, and calibration monitoring
- LIMS-compatible import + export with customizable report generation
- Refined algorithms for peak detection, baseline handling, and carry-over correction
- System configuration and method parameters are archived with data in an individual result file
- Unparalleled access to instrument component diagnostics in real-time, even during data collection
Methods*
Analyte | Technique | Method | Operating Range | MDL | Throughput |
---|---|---|---|---|---|
Ammonia Nitrogen | FIA | USEPA Method
350.1 |
0.01-20 mg/L ammonia as nitrogen |
0.002 mg/L | 30 samples per hour |
Available Cyanide | FIA | USEPA Method
OIA-1677-09 |
2.0 μg/L–5.0 mg/L | 0.5 μg/L | 30 samples per hour |
Nitrate/Nitrite | FIA
SFA |
USEPA Method
353.2 |
0.01–10.0 ppm 10.0-10,000 ppb0.005-10.0 ppm |
0.001 ppm 1.0 ppb0.001 ppm |
60 samples per hour40 samples per hour |
Nitrogen
(Total Dissolved) |
SFA | USEPA Method 353.2 |
0.01 – 20 mg/L | 0.006 mg/L N as NO3 – |
20 samples per hour |
Phenol
In-Line Distillation |
SFA | USEPA Method 420.4 |
1.0-500 ppb | 0.5 ppb | 22 samples per hour |
Phenol
(Post-Distillation) |
FIA | USEPA Method 420.4 |
0.01-2.00 ppm 10.0–2,000 ppb |
0.002 ppm 2.0 ppb |
90 samples per hour |
Phosphorus
(All Forms) |
FIA | USEPA Method 365.1 |
0.01–5.0 ppm 10.0-5,000 ppb |
0.001 ppm 1.0 ppb |
60 samples per hour |
Phosphorus
All Forms- Low Level |
FIA | USEPA Method 365.1 |
0.001-0.1 ppm 1.0-100 ppb |
0.0003 ppm 0.3 ppb |
45 samples per hour |
Phosporus(Total) | FIA w/ In-line UV/Persulfate | Standard Method 4500-P-I |
0.010 mg/L – 10 mg/L P | 0.002 mg/L | 30 samples per hour |
Silica | SFA | Standard Method 4500-SiO2 E |
0.02 – 20 mg/L SiO2 | 0.02 mg/L | 60 samples per hour |
Sulfate | FIA
Photometric |
USEPA Method 375.2 |
1.0 mg/L – 25 mg/L |
0.1 mg/L | 40 samples per hour |
TKN
(Total Kjeldahl Nitrogen) |
SFA, Gas Diffusion | USEPA Method 351.2 |
0.01-20.0 ppm 10.0-20,000 ppb |
0.001 ppm 1.0 ppb |
40 samples per hour |
*More methods available, these are the most popular options
Technical Specifications
Analysis Module: 1 or 2 chemical analysis channels per chassis
Analysis Module Dimensions: 31 in. W x 17.5 in. D x 10.5 in. H; 78.74 cm W x 44.45 cm D x 26.67 cm H
FS 3700 Dimensions with 90-position Autosampler: Approximately 44 in. (112 cm) W
FS 3700 Dimensions with 360-position Autosampler: Approximately 51.5 in. (131 cm) W
Injection Valve: 8 or 10-port switching valve with chemically-inert wetted surfaces
Photometric Detector: 420-880 nm, with PEEK path lengths of 5-, 10- or 20-mm
Amperometric Detector: Silver working electrode; silver/silver chloride reference; stainless steel counter electrode
In-line Heater: Included as needed, mounted underneath chemistry cartridge user programmable in 1 ̊C increments
UV-digestion Module: Included as needed, mounted underneath chemistry cartridge
Peristaltic Pump: 24-channel, fits on top of analysis module
Autosampler: 90-position, X-Y-Z (90 samples + 9 standards); 360-position, X-Y-Z (360 samples + 10 standards)
Tubing: FEP Teflon® and EVA ethylene-vinyl acetate copolymer
Manifolds / Fittings: Polysulfone
Analysis Methods / Documentation: Validated chemistries for specific analytes / sample matrices with performance data
Operating Software: FlowView
Operating System: Windows® 7, Windows® 8, 8.1 and 10
Data Collection: 6 channels per instance of software; Multiple instances of software can be run on a single computer
PC to FS 3700 Communications: USB
Power Supply: 24VDC universal switching power supply for operation with 90-250VAC 50/60Hz source
Power Requirements: 110VAC/60 Hz or 230VAC/50 Hz
Weight (Analysis Module): 19.5 kg (43 lbs.), typical for analysis module and pump, two injection valves, chemistry cartridges, detector modules
Certifications: CE Safety EN 61010-1; EMC Immunity & Emissions EN 61326-1:2006